AIR X SOFTWARE TUTORIAL

DEVICE VIDEOS

BreatheLITE lets you seamlessly import MP4 videos from video cameras and view them overlaid with synchronised dust monitor data. You can pinpoint when and where particle spikes occur, inspect timestamps and sensor readings together. 

You can only import .mp4 files up to 2GB in size. If you have a video that is not in this format, you will need to use a free online converter to change it into the correct format. 

 

BreatheLITE Device Videos tutorial

ADDING A VIDEO

To add a video, click the Device Button and select Videos. The video library dialog will open. In the Import Video tab, Click the Browse button and an open file dialog will be displayed. Select the desired .mp4 video file to import it. The file will then begin to be imported into BreatheLITE. 

Once the video is uploaded to the software, the next step is to confirm whether the timestamp of the video is correct. Getting the correct timestamp is vital for BreatheLITE to correctly overlay the device readings when a video is being viewed. You can amend any times or dates of videos by pressing the pencil button and click Save. The video will then import. 

A confirmation dialog will then be displayed to confirm the import. Click on the Yes button to complete the video import process. The video will then be available in the video library. 

BreatheLITE Device Videos tutorial: Adding a video

VIEWING A VIDEO USING VIDEO LIBRARY

To view a video using the Video Library, click the Device Button and select Videos. The video library dialog will open. Click on the Video Library tab and select the video you wish to view by clicking on the Play button. The video will then be displayed. 

To view a video using the Data Analysis component, select a device, date, PM size, Units of Measurement and all relevant Display Options and view the data. Once the chart has been displayed, if a video is found, a green play icon will appear on the chart. To view the video, click on the green play icon and the video will be displayed. 

BreatheLITE Device Videos tutorial: Viewing a video

VIEWING A VIDEO USING DATA ANALYSIS

To view a video using the Data Analysis component, select a device, date, PM size, Units of Measurement and all relevant Display Options and view the data. Once the chart has been displayed, if a video is found, a green play icon will appear on the chart. To view the video, click on the green play icon and the video will be displayed. 

BreatheLITE Device Videos tutorial: Using video for data analysis

DELETING A VIDEO

To delete a video, click the Device Button and select Videos. The video library dialog will open. Click the Video Library tab and click the Delete button by the video you want to delete. A confirmation dialog will then be displayed. To delete the video, click the Yes button and the video will be removed. 

BreatheLITE Device Videos tutorial: Deleting a video

GOT PRODUCT QUESTIONS?

If the Trolex XD1+ STEL and TWA indicators are flashing red as shown below:

This means the Trolex XD1+ has detected an electronic failure of the internal memory. The XD1+ will alert the user and enter an error state.

If this happens, please contact our support team to raise a ticket.

AIR XD operates a protective thermal cut-out sequence when the temperature inside the instrument exceeds the maximum operating temperature of 45 °C.

This protective measure is in place to maintain the lifespan and operating functionality of the optical sensor assembly when the AIR XD is installed in environments with high ambient temperatures.

If the internal temperature of the instrument is suspected to be below -10°C, for example, if it has been left powered off for a period of time in a very cold environment, damage to the optical assembly may occur when power is applied.

Allow the instrument to reach safe operating temperature before applying power.

If all three device indicators are flashing red as shown below, this means the device has a non-recoverable operating error. 

If this is happening to your Trolex XD1+ device, please contact the Trolex service team to raise a new support ticket

Our AIR X Particulate Monitors don’t hold any of these accreditations.

No. The Group I manual states that the USB connector on the base of the equipment shall only be connected to the Trolex TX8805 AIR XD USB Drive in the hazardous area.

USB drives with no power source, such as standard USB sticks, may be connected to the USB connector in the safe area for test purposes.

AIR XD is designed to allow infrequent pressure washing during maintenance schedules, but the inlet cap must be closed first.

Keep an appropriate distance between the pressure washer and the instrument, and do not direct the water jet at the underside of the instrument because water must not enter the air flow path.

Yes. The AIR XD provides in and out RS485 terminals so multiple MODBUS devices can be daisy-chained.

Some network installations may require a termination resistor across RS485 lines A and B.

SEE MORE RESOURCES

AIR X PARTICULATE MONITORING

Image of TX8061 XD1+ Personal Dust Monitor
TX8061 XD1+ Personal Dust Monitor

Image of AIR X Software
AIR X Software

Image of TX8005 AIR XD Dust Monitor
TX8005 AIR XD Dust Monitor

Image of TX8100 AIR XS Silica Monitor
TX8100 AIR XS Silica Monitor

AIR X NEWS + CASE STUDIES

GET AIR X NEWS + UPDATES

    You may unsubscribe from these communications at any time. For more information on how to unsubscribe, our privacy practices, and how we are committed to protecting and respecting your privacy, please review our Privacy Policy.