Pellistor Protect is deployed when measured gas values exceed 100% of full scale on a pellistor sensor.

The display shows Pellistor Lockout alarm, the analogue output is clamped at 22 mA or 2.2 V, MODBUS gas value reads -10,000 with the over-range flag set, and assigned relays or A/V alarms activate.