AIR XS SILICA MONITOR

Real-time silica monitoring for workers at risk from respirable crystalline silica.

AIR XS SILICA MONITOR

Real-time silica monitoring for workers at risk from respirable crystalline silica.

  • Real-time, continuous measurement of atmospheric RCS concentration
  • High-reliability and low-maintenance
  • On-device display readout
  • High visibility alarm warning indicators
  • ‘Plug and play’ installation
  • Patented Optical Refraction Technology (ORT)
  • Real-time data with instant visualisation
  • Almost zero maintenance
  • No complicated start-up routines
  • Detects all forms of crystalline silica and can track changing levels of RCS in complex dust mixtures over time. No more guess work; know exactly how much RCS dust your workers are exposed to in real-time.
  • Monitor for up to 24 hours a day, with no complicated set-up, no calibration routines, no return-to-base servicing, no moving parts and no fuss. Work as usual without interruption or hassle whilst knowing the exact exposure to dangerous dust.
  • Instant alarm allows for decisions to be made regarding use of PPE and RPE, agile process development and smart deployment of ventilation and extraction systems, reducing costs, improving business efficiency, and ultimately saving lives.
  • Ruggedised design, IP66 ingress protection and additional transportable pack means that the AIR XS can be located in any area across your workplace, designed to function in any challenging industrial environment.
  • Seamless IoT integration allowing for real-time and remote management of your entire particulate monitoring processes.

      GOT PRODUCT RELATED QUESTIONS?

      Got a question relating to our AIR XS Silica Monitor? Take a look at our FAQs. If you can't find what you're looking for then you can submit a support ticket and an expert will be in touch.

      Yes. The manual states that AIR XS may show cross-responsivity from poly-crystalline particles in the local environment, with severity depending on particle density and crystalline properties.

      Because of these possible interferences, the manual frames AIR XS as an indicative real-time monitoring device rather than a regulatory compliance method.

      Yes, via MODBUS RTU (RS485) or MODBUS TCP/IP (Ethernet) for integration with SCADA systems and Aethair’s Thiamis.

      AIR XS has two user-configurable alarms for warning of a selected particulate threshold breach. From Main Menu, select Alarms, use Detail to view current alarms, and Configure to set channel, threshold value, latching or non-latching mode, and units.

      To save changes, exit using the left arrow key and confirm the Settings saved dialogue.

      1. Press and hold the centre button to go to the Main Menu
      2. Go to Communication.
      3. Configure RS485 baud rate and Ethernet settings.

      Open Main Menu > Communication. Use the RS485 or Ethernet submenu to view settings through Details or change them through Configure.

      If floating point byte order is not compatible with the receiving system, change the byte ordering from Main Menu > Communication > RS485/MODBUS > Configure.

      1. Go to Main Menu > Alarms
      2. Choose a measurement period (1 min, 15 min, 1 hr, etc.)
      3. Set a threshold limit
      4. Select alarm mode (latching or non-latching)
      1. Insert a USB device (minimum 8GB, FAT32 format)
      2. Press and hold the centre button to go to the Main Menu
      3. Navigate to Data Download > USB Download

      PARTICULATE MONITORING

      Image of TX8061 XD1+ Personal Dust Monitor
      TX8061 XD1+ Personal Dust Monitor

      Image of TX8005 AIR XD Dust Monitor
      TX8005 AIR XD Dust Monitor

      Image of TX8060 XD ONE Portable Dust Monitor
      TX8060 XD ONE Portable Dust Monitor

      Image of AIR X Software
      AIR X Software

      SALES ENQUIRIES

      GET PRODUCT UPDATES

        You may unsubscribe from these communications at any time. For more information on how to unsubscribe, our privacy practices, and how we are committed to protecting and respecting your privacy, please review our Privacy Policy.