- Select the appropriate Channel
AIR XD has 2 x alarm channels that can be assigned one of four particulate sizes or TSP. To select the appropriate channel, go to Settings > Alarms > Alarm 1 > Channel Select and select the PM size that you want to monitor. - Select the Threshold Channel
Select and assign an alarm to a specified data source. To select the data source, go to Settings > Alarms > Alarm 1 > Channel Select and select the PM size that you want to monitor.
- Set Alarm Threshold
Set the threshold value, which activates the alarm. Threshold values can be configured between 1 μg/m3 to 9,999 μg/m3 or 0.1 mg/m3 to 6,553.5 mg/m3. To set the threshold, go to Settings > Alarms > Alarm 1 > Set Threshold and select enter the value of the threshold.
- Set Alarm Latching
Configure the latching behaviour for the selected alarm. To set the alarm latching, go to Settings > Alarms > Alarm 1 > Latching En/Dis and select whether you want to enable or disable latching. -
Enable/Disable the Alarm