Navigate to Settings > Maintenance > Compliance Test. The instrument resets the memory module, prompts cleaning, asks the user to confirm the top cap is open, and then prompts fitting the particulate dispersion hood.

Dose the hood with reference material during the first 10 seconds of the test. The AIR XD then analyses the sample and returns a pass or fail result.