Confirm the correct zero gas and calibration gas are being used for the fitted gBloc, that the gas flow regulator is set to 0.5 L/min to 1 L/min, and that the gas reading has stabilised before adjustment.

Remember that all gas sources have error tolerances and calibration must be performed by competent accredited personnel.